_________Dialysis You Can Trust

 

Technology

A dialysis system with cutting edge technology:
· Full touch screen showing all data in a single glance
· Redundant system to insure maximum Safety
· RS-232 port for maintenance at a distance
· And yet very user friendly

High precision & safe UF control & measurements Technology using a true single pass for with error as low as 1g /hr.

Safety is insured by various programmable disinfection techniques:
· Chemical cleaning
· Chemical disinfection
· Thermo Chemical disinfection
· Thermal disinfection



Profiling

· Sodium
· Bicarbonate
· Ultrafiltration
· Heparin
· Hemodiafiltration HDF
· HDF w/reinfusion
· PHF online HDF


Accessories & options

· Water ultra filter
· Dialysate ultrafilter
· Sphygmomanometer
· Caedic function monitor
· Hematocrit monitor


Treatment Modality

· Double needle dialysis
· Single needle dialysis w/ single pump
· w/ double pump
· Sequential dialysis
· Acetate dialysis
· Bicarbonate dialysis
· Liquid Bicarbonate Unidose

 

 

 

 
Technical Data

· Fully computerized system
· Full touch screen software
· RS 232 integrated opto-isolated connection for maintenance

Blood flow control: 20 ml/min steps
Single pump 0 to 800 ml/min
Double pump 0 to 600 ml/min

Blood pressure:
Venous -400 to 880 mm Hg
Arterial -400 to 880 mm Hg
TMP -60 to 520 mm Hg

Heparin pump:
Syringe size 20, 30 & 50 ml
Flow rate 0.1 to 10 ml/hr

Alarm systems:
Air bubbles on venous line
Technique Ultrasound
Sensitivity bubbles ≥ 10 µl

Blood on arterial & venous lines:
Technique IR light
Blood leaks IR light

Dialysis Fluids:
Flow rate 300 to 800 ml/min
Temperature 35°C to 39°C
Ultrafilter MF-100-DC

Conductivity:
Total 12.1 to 15.8 mS/cm
First step 2.4 to 3.6 mS/cm
4.0 to 6.0 mS/cm

Ultra filtration control:
Absolute Single Pass Coriolis Force
UF rate 0.1 to 2.5 kg/hr
Accuracy ± 1 g/min or ± 1%

Dimensions & requirements:
Size HxWxD 1740 x 500 x 755 mm
Weight 98 kgs
Water quality > AAMI norms
Water Temperature °C to 39
Filter MF 900

Power requirements:
Voltage 115–230 VAC ± 10%
Frequency 50 to 60 Hz

 
   
 
Copyright 2012 Meditechlab